TECHNICAL DETAIL
Double Beam Optics
Wavelength Range: 190 - 1100 nm or better Bandwidth : 0.5 to 20 nm Variable Band Width Wavelength accuracy: +/-0.1nm
Wavelength reproducibility: +/-0.1nm
Concave holographic grating with minimum 1050 lines/nm or better Stray Light: 0.02%T at 220nm (NaI) or better.
Photometric Accuracy: ± 0.004A Photometric reproducibility: <0.1nm
Photometric noise at 700nm (0A) must be less than 0.00005A or better Pre-aligned Deuterium and Tungsten - Halogen lamps with automatic switch-over/xenon flash
PC controlled software would include modules for Scan, Time Drive,
Wavelength Programming, Concentration, Bio Assay, Kinetics, Oligo
Quant, Validation, Report Builder, Arithmetic etc. as standard features of the s/w.
The manufacturer would be ISO-9001 Certified Company. Accessories: Suitable PC, Printer, Suitable Online UPS, Cuvette: 10mm path-length, 3.5 ml Appx (02 Nos)
Note: Equipment would be complete with all mandatory accessories and ready to use manner with guide manual. |