Description:
Defibrillator is required for reviving the heart functions by providing selected quantum of electrical shocks with facility for monitoring vital parameters.
Operational Requirements:
Defibrillator is Bi-Phasic, Light weight and latest model
Monitor vital parameters and display them
Print the ECG on thermal recorders.
Should work on both Manual and Automated external defibrillation (AED) mode up to 200 j or more.
Capable of doing synchronized & a synchronized cardio version
Can be operated from mains as well as battery
Defibrillator testing facility
Technical Specifications:
Low Energy Biphasic defibrillator monitor with Recorder, having capability to arrest all arrhythmia within a maximum energy of 200 joules.
Monitor ECG through paddles, pads and monitoring electrodes and Defibrillate through pade and paddles. Automatic or Manual Lead switching to see patient ECG through paddles or leads
Measure and compensate for chest impedance for a range of 25 to 125 ohms
Built in 50mm strip Printer/ Thermal recorder
Charging time of less than 6 for maximum energy. Charging indicator is there.
Bright LCD / TFT display for viewing messages and ECG wave form of 4 seconds.
|