TECHNICAL DETAIL
ASTM D86 (group 0 to 4), ASTM D850, ASTM D1078, ISO 3405, IP123, IP 195
Distillation rate: 2 to 15 ml/minutes in 0.1ml increments
Pt-100 vapor temperature sensor with automatic temperature calibration system (°C or °F)
Temperature - Distillation Range: 0 to 450°C (±0.1°C accuracy) ;
Condenser: -5 to 60°C (±0.1°C accuracy) closed loop system ;
Receiver Chamber: 0 to 60°C (±0.1°C accuracy)
Optical measuring system compatible with samples producing smoke in the receiver; range 0 to 103% charge volume; resolution: 0.03ml, accuracy: ± 0.1ml
Built-in pressure sensor, range to 70 to 110 kPa (ab) (500 to 800 mmHg) Calibration: Single point against reference barometer
Automatic determination of initial boiling point (IBP), final boiling point (FPB), dry point and barometric and residue corrections
Automatic self check at instrument start
Sample Heater: Automatic heater adjustment for last 5 ml, Automatic heater lowering at the end of test, Automatic cooling of distillation flask to keep minimum test interval time
Distillation of unknown samples: distill unknown samples precisely in the very first run without need to run several trials.
Automatic and precise calculation of residue loss to avoid error on residue measurement.
Test results must be displayed in real-time and must include distillation curve and temperature with barometric compensation and evaporation correction, distillation rate, heating power curve, master curve comparison and zoom function for high resolution of heating and temperature curves.
Must have Built-in safety device with the system. Automatic safety shut down of heater and other power devices.Enlist all Error message for easy service diagnostics and quick rectification.
Transparent door for viewing of test samples in the receiver chamber. Multi-level access control for running tests, defining unit configuration and maintenance (calibration & diagnostic).Automatic exhaust of vapour from receiver chamber at the end of test to reduce VOC emission in the laboratory.
Equipment capable for Auto calibration of temperature, volume and pressure and automatic recording and storage of all calibrations for paperless tracking.
Necessary Certified Reference Standards with traceability certificate to be supplied.
Colour touch screen: Built-in integrated large LCD display for displaying distillation curves and reports.
The analyzer output for connecting external printer.
The apparatus provision to connect several distillation units for control through external PC and software (ready for future expansion).
The system at least 2 USB for external printer, memory stick, RS 232 C serial port for networking and other purposes.
Device integrated computer with device software.
EXTENDED DISTILLATION:
Precise distillation rate regulation for ethanol blended samples (E10, E15 etc.)
Have Provision to distill in manual mode.
Essential Consumable parts to be supplied for distillation unit
I) Distillation Flasks D86: 125ml capacity: 10 no
II) Receiver Glass Cylinder: 10 no
III) Distillation Probe: 2 no
IV) Heater plate, Ceramic x 2 each of 32 mm, 38 mm and 50 mm dia.
V) Fuses: 1 set
VI) Silicone hose for Flask connection to condenser tube: 10
VII) Certified Reference Material for Distillation – 250ml x 2 bottles
VIII) O-ring for vapor seal: 6
IX) Other critical spares which will be required at the time of commissioning. |